The Solar 9354-2 is a transient generator for military and aerospace equipment testing, complying with MIL-STD-461G and RTCA DO160D standards. This model features nine waveforms, a digital voltmeter, and adjustable output voltage, suitable for various susceptibility tests.
Features
- Panel-mounted digital voltmeter. Monitors the adjusted open circuit discharge voltage
- Pulse rates up to one pulse per second maximum (factory adjusted)
- Single pulse feature enables controlled isolation of transient effects
- Output voltage adjustable from 0.1% to 100% of selected discharge voltage
Specifications
Damped Sinusoid Pulses
|
10 KHz |
100 KHz |
1 MHz |
10 MHz |
30 MHz |
100 MHz |
Open Circuit Voltage |
30 V |
300 V |
3200 V |
3200 V |
1000 V |
300 V |
Calculated Short Circuit Current |
120 A |
120 A |
128 A |
128 A |
20 A |
6 A |
Source Impedance |
<0.25 ohms |
<2.5 ohms |
<25 ohms |
<25 ohms |
<50 ohms |
<50 ohms |
Double Exponential Pulses
|
6.4 μS |
70.0 μS |
120 μS (9354-2) |
Rise Time |
100 nS |
6.4 μS |
40 μS |
Open Circuit Voltage |
1600 V |
1600 V |
1600 V |
Calculated Short Circuit Current |
800 A |
800 A |
400 A |
Source Impedance |
<2.0 ohms |
<2.0 ohms |
<4.0 ohms |
Accessories
Type 9335-2
Universal Coupling Device |
An inductive injection device that provides voltage and current transfer of 1:1, 1:1.5 and 3:1 voltage
step-up (current step-down) as well as 2:1 voltage
step-down (current step-up) |
Type 9719-1N
Injection Probe |
Provides the required current levels of MIL-STD-461D and E, method CS116 throughout the entire frequency
range of 10 KHz to 100 MHz |
Type 9357-1
Calibration Fixture |
Calibration fixture provides a 50 ohm characteristic
impedance based on the dimensions of the Type 9335-2 Universal Coupling Device and Type 9719-1N Injection Probe. The fixture maintains a low standing wave ratio from 10 KHz to 100 MHz in a 50 ohm circuit. |
Type 9142-1N
Injection Probe |
Used to inject current on cables from 1 MHz to 100 MHz. |
Type 9125-1
Calibration Fixture |
Calibration fixture for use with Type 9142-1N Injection
Probe |
Type 9123-1N
Current Probe |
Used to monitor injected pulses. Frequency range from 10 KHz to 500 MHz |
Type 9410-1
50Ω/50Ω and
Type 9454-1
600Ω/50Ω
High Voltage Attenuators |
Provides 40 dB attenuation from 10 KHz to 100 MHz. Protects oscilloscope from high voltage damage when verifying the output pulses of the Model 9354-2. The Type 9454-1 provides a high impedance to the oscilloscope for making measurements of open circuit output pulses |
Type 6220-4
High Voltage Audio Isolation Transformer |
When connected in series with the power lead under test, provides twice the open circuit voltage or twice the short circuit current for the 10 KHz and 100 MHz damped sinusoid waves and the 6.4 S and 70 S double exponential pulses. Capable of handling up to 4000 volts |
Type 9616-1
Injection Clamp |
Provides coupling for high voltage pulses produced by the Model 9354-2. Meets the inductive indirect injection device requirement of MIL-STD-462, notice 5, method CS10 and CS11. Enables injection of 70 S double exponential pulses without need for direct connection. |