- 100 watt source or sink
- < 1ms transient recovery
- Built-in voltage and current monitors
As a Constant Voltage Source
Output | Compliance (Source or Sink) |
Range | Maximum | Resolution | Maximum | Resolution |
100 V | ± 101.0 V | 100 mV | ± 1.010 A | 1 mA |
± .1010 A | 100 µA |
10 V | ± 10.10 A | 10 mV | ± 10.10 A | 10 mA |
± 1.010 A | 1 mA |
± .1010 A | 100 µA |
1 V | ± 1.010 A | 1 mA | ± 10.10 A | 10 mA |
± 1.010 A | 1 mA |
± .1010 A | 100 µA |
- Output Resistance (maximum)
- 100 V Range: 10 m ohms.
- 10 V Range: 100 µ ohms.
- 1 V Range: 100 µ ohms.
- Output Inductance - 100 µ H typical
- Sensing - REMOTE or LOCAL
As a Constant Current Source
Output | Compliance (Source or Sink) |
Range | Maximum | Resolution | Maximum | Resolution |
10 A | ± 10.10 A | 10 mA | ± 10.10 V | 10 mV |
± 1.010 V | 1 mV |
1 A | ± 1.010 A | 1 mA | ± 101.0 V | 100 mV |
± 10.10 V | 10 mV |
± 1.010 V | 1 mV |
0.1 A | ± 0.1010 A | 100 µA | ± 101.0 V | 100 mV |
± 10.10 V | 10 mV |
± 1.010 V | 1 mV |
- Output Resistance (min)
- 10 A Range: 104 ohms
- 1 A Range: 10 5 ohms
- 0.1 A Range: 106 ohms
- Output Capacitance - 0.2 µ F typical
- Output Load - Must be non-inductive
Current Monitor Output
- Scale Factor - 1 V = 100% of range
-
- External Modulation
- Sensitivity - -10V increases magnitude of programmed output by 100% full scales; + 10 V decreases magnitude of programmed output by 100% of full scale
- Modulation Frequency - 600 Hz bandwidth
-
- General
- Display - Dual 3½-digit LED (0.5 in) displays indicate programmed values in Standby and output values in Operate
- Program Memory (battery backed-up) - Stores up to 100 output settings.
- Range of Dwell Times: 10 ms to 1000 s
- Output Connections - Quick disconnect card with screw terminals for output, modulation, current monitor, and external sense. BNC (chassis isolated) connectors for TRIGGER IN/OUT.