Com-Power
Com-Power CDN-S1 Coupling/Decoupling Network (CDN)
The Com-Power CDN-S1 Coupling/Decoupling Network (CDN) is for testing from 150 kHz - 80 MHz according to the EN 61000-4-6 / IEC 61000-4-6 standard for immunity to conducted disturbance induced by radiofrequency fields. The Com-Power CDN-S1 has both coupling and decoupling networks. The coupling network delivers injected RF common mode current disturbance signals through the various conductors to the equipment under test (EUT).
The decoupling networks are used to ensure that the disturbing signals injected on the signal or power line of the EUT by the coupling networks do not interfere with any of the auxiliary equipment (AE) connected to the EUT. The Com-Power CDN-S1 contains integrated direct capacitive coupling along with a high impedance choke for inductive decoupling. The disturbing signal can be injected on signal lines and power lines of a EUT by using the appropriate CDN described below. The CDN setup must be calibrated to establish the drive table. Please use the appropriate calibration components for the CDN
For each CDN you will need:
- 2 units of the 150 Ω to 50 Ω adapter
- 1 unit of the 50 Ω Terminator
- Optional Ground plane
- Calibration adapters
S Series CDNs for shielded signal cables
The S series CDNs are for shielded cables with single or multiple conductors. These CDNs are supplied with professional audio, D type and Centronics connectors.
# wires in the cable |
1 |
Coupling Point |
Shield |