Warm-up Time |
Twice the time since last warmed up, to a maximum of 30 minutes |
Settling Time |
5 seconds or faster for all functions and ranges |
Standard Interfaces |
IEE-488 (GPIB), RS-232 |
Temperature Performance |
Operating: 0 °C to 50 °C
Calibration (tcal): 15 °C to 35 °C
Storage: -20 °C to 70 °C |
Electromagnetic Compatibility |
Designed to operate in Standard Laboratory environments where the Electromagnetic environment is highly controlled. If used in areas with Electromagnetic fields > 1 V/m, there could be errors in current output values. |
Temperature Coefficient |
Temperature Coefficient for temperatures outside tcal ±5°C: add 0.1 x 1-year specification/°C |
Relative Humidity |
Operating: <80% to 30 °C, <70% to 40 °C, <40% to 50 °C
Storage: <95%, non-condensing |
Altitude |
Operating: 3,050m (10,000 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum |
Safety |
Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010, 1-92 |
Analog Low Isolation |
20V |
EMC |
Complies with EN 61326-1 |
Line Power |
Line Voltage (selectable): 100V, 120V, 220V, 240V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: ± 10% about line voltage setting |
Power Consumption |
250 VA |
Dimensions |
Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
Width: 43.2 cm (17 in), standard rack width
Depth: 47.3 cm (18.6 in) overall |
Weight |
20 kg (44 lb) |