Temperature Range |
-73ºC (-100ºF) to 190ºC (375ºF) Cascade
-34ºC (-30ºF) to 190ºC (375ºF) Single Stage |
Temperature Stability |
±1ºC (±2ºF); ±4% relative humidity at steady state condition after stabilization (usually 1 hour)
±1ºC (±2ºF); at steady state condition after stabilization (usually 1 hour) |
Temperature Performance |
Cascade:
Temperature pulldown from 23ºC (75ºF) to:
-40ºC (-40ºF) in 20 minutes.
-54ºC (-65ºF) in 30 minutes.
-68ºC (-90ºF) in 45 minutes.
Single Stage:
Temperatue pulldown from 23ºC (75ºF) to:
-18ºC (0ºF) in 15 minutes.
-34ºC (-30ºF) in 30 minutes.
Temperature heatup from 23ºC (75ºF) to:
+93ºC (200ºF) in 20 minutes. | +190ºC (375ºF) in 55 minutes. |
Live Load Capacity - Cascade |
1500 watts at -40ºC (-40ºF)
1100 watts at -54ºC (-65ºF) |
Live Load Capacity - Single Stage |
1700 watts at -18ºC (0ºF)
760 watts at -34ºC (-30ºF) |
Interior Dimensions |
965mm (38in)W x 965mm (38in)D x 965mm (38in)H |
Exterior Dimensions |
1320mm (53in)W x 1630mm (66in)D x 1960mm (78in)H |
Interior Dimensions |
965mm (38in)W x 965mm (38in)D x 965mm (38in)H |
Exterior Dimensions |
1320mm (53in)W x 1630mm (66in)D x 1960mm (78in)H |
Shipping Weight - Cascade |
Approx. 2175 lbs |
Shipping Weight - Single Stage |
Approx. 2075 lbs. |
Utility Requirements: 208/230 volt (-0%, +5%), 60 Hz. |
1 phase: 51 full load amperes.
3 phase: 39 full load amperes. |
Heat Rejection |
Approximately 9,000 watts (30,000 BTU/hr) maximum during cooling operation. |