Magnetic (B-) Field |
Sensor System |
Coil (100 cm2) |
Measurement Axis |
Tri-Axial (isotrop) |
Exposure STD Mode |
Frequency Range (-3dB), selectable |
1 Hz to 400 kHz, 10 Hz to 400 kHz, 30 Hz to 400 kHz |
Exposure Evaluation |
Compared to Standard |
MODE |
ICNIRP |
BGV B11 |
EN 50366 |
RANGE |
LOW |
HIGH |
LOW |
HIGH |
LOW |
HIGH |
Overload Limit |
160% |
1.600% |
160% |
1.600% |
160% |
1.600% |
Nominal Range |
10% |
100% |
10% |
100% |
10% |
100% |
Noise Level, approximate |
1.3% |
6% |
0.4% |
6% |
0.4% |
6% |
Damage Level RMS (Peak) |
160 mT (226 mT) |
Resolution (RANGE: LOW) |
0.001 % |
Detection, selectable |
Automatic according to selected Standard or RMS (averaging time 1 s) or Peak Value |
Field Strength Mode |
Frequency Range (-3dB), selectable |
1 Hz to 400 kHz, 10 Hz to 400 kHz, 30 Hz bis 400 kHz |
Frequency Response |
Flat |
MODE |
320 µT |
8 mT |
80 mT |
RANGE |
LOW |
HIGH |
LOW |
HIGH |
LOW |
HIGH |
Overload Limit |
32 µT |
320 µT |
800 µT |
8 mT |
8 mT |
80 mT |
Nominal Range |
2 µT |
20 µT |
50 µT |
500 µT |
500 µT |
5 mT |
Noise Level, approximate |
70 nT |
320 nT |
1.8 µT |
8 µT |
18 µT |
80 µT |
Damage Level RMS (Peak) |
160 mT (226 mT) |
Resolution (RANGE: LOW) |
1 nT |
Overall Uncertainty |
±4 % (50 Hz to 120 kHz) |
Detection, selectable |
RMS (averaging time 1 s) or Peak Value |
Display Mode, selectable |
Instantaneous or Max Hold |
Output |
Analog Scope Output |
Three Channel (X-Y-Z) |
Output Level |
50 mV rms at scale final value of Nominal Range (Impedance: 50 Ohm, nominal) |
Interface (Remote Control and Readout) |
RS-232 (19200 baud, 8n1, XON/XOFF), 3-Wire, Stereo Jack 2.5 mm |